1.1 This practice provides a procedure for computing a 99 %/95 % Within-laboratory Detection Estimate (WDE) and the associated critical level/value (WCL). The WDE is the minimum concentration, with false positives and false negatives appropriately controlled, such that values above these minimums are reliable detections. The WCL is the point at which only false positives are controlled appropriately. A false positive is the reporting of an analyte as present when the analyte is not actually present; false negatives are reports of analyte absence when the analyte is actually present. This practice is distinguished from the Interlaboratory Detection Estimate (IDE) practice in that the IDE Standard utilizes data from multiple, independent laboratories, while this practice is for use by a single laboratory. The IDE would be utilized where interlaboratory issues are of concern (for example, limits for published methods); this practice (and values derived from it) are applicable where the results from a single laboratory, single operator, single instrument, etc. are involved (for example, in understanding, censoring and reporting data).
1.2 The establishment of a WDE involves determining the concentration below which the precision and bias of an analytical procedure indicates insufficient confidence in false-positive and false-negative control to assert detection of the analyte in the future analysis of an unknown number of samples. Most traditional approaches attempt to determine this detection “limit” by estimating precision at only a single, arbitrary point. The WDE approach is intended to be a more technically rigorous replacement for other approaches for estimating detection limits. The WDE practice addresses a number of critical issues that are ignored in other approaches.
1.3 To summarize, the WDE is computed to be the lowest true concentration at which there is 90 % confidence that a single (future) measurement (from the studied laboratory) will have a true detection probability of at least 95 % and a true non-detection probability of at least 99 % (when measuring a blank sample). For the laboratory in the study, the critical value is the true concentration at which, on average, (with approximately 90 % confidence) will not be exceeded by 99 % of all measurements of samples with true concentration of zero (that is, blanks). These values are established by modeling the precision and establishing the recovery/bias over a range of concentrations, as well as by using a tolerance interval. The complexities of the WDE procedure may appear daunting, but the additional considerations are necessary if meaningful estimates of the actual detection capabilities of analytical methods are to be made. The concepts are tractable by degreed chemists, and the use of the available ASTM DQCALC Excel-based software makes the data analysis and limit determinations easy.
1.4 A within-laboratory detection estimate is useful in characterizing the concentration below which a method, for an analyte, as implemented in a specific laboratory, does not (with high confidence) discriminate the presence of the analyte from that of the absence of an analyte. As such an estimator, the WDE Standard (and the WDE and WCL values produced through its application) are useful where a trace-analysis testing method needs to be used.
1.1 This practice provides a procedure for computing a 99 %/95 % Within-laboratory Detection Estimate (WDE) and the associated critical level/value (WCL). The WDE is the minimum concentration, with false positives and false negative appropriately controlled, such that values above these minimums are reliable detections. The WCL is the point at which only false positives are controlled appropriately. A false positive is the reporting of an analyte as present when the analyte is not actually present; false negatives are reports of analyte absence when the analyte is actually present.
This practice is distinguished from the Interlaboratory Detection Estimate (IDE) practice in that the IDE Standard utilizes data from multiple, independent laboratories, while this practice is for use by a single laboratory. The IDE would be utilized where interlaboratory issues are of concern (for example, limits for published methods); this practice (and values derived from it) are applicable where the results from a single laboratory, single operator, single instrument, etc. are involved (for example, in understanding, censoring and reporting data).
1.2 The establishment of a WDE involves determining the concentration below which the precision and bias of an analytical procedure indicates insufficient confidence in false-positive and false-negative control to assert detection of the analyte in the future analysis of an unknown number of samples. Most traditional approaches attempt to determine this detection “limit” by estimating precision at only a single, arbitrary point. The WDE approach is intended to be a more technically rigorous replacement for other approaches for estimating detection limits. The WDE practice addresses a number of critical issues that are ignored in other approaches.
1.3 To summarize, the WDE is computed to be the lowest true concentration at which there is 90 % confidence that a single (future) measurement (from the studied laboratory) will have a true detection probability of at least 95 % and a true non-detection probability of at least 99 % (when measuring a blank sample). For the laboratory in the study, the critical value is the true concentration at which, on average, (with approximately 90 % confidence) will not be exceeded by 99 % of all measurements of samples with true concentration of zero (that is, blanks).
These values are established by modeling the precision and establishing the recovery/bias over a range of concentrations, as well as by using a tolerance interval. The complexities of the WDE procedure may appear daunting, but the additional considerations are necessary if meaningfully estimates of the actual detection capabilities of analytical methods are to be made. The concepts are tractable by degreed chemists, and the use of the available ASTM DQCALC Excel-based software makes the data analysis and limit determinations easy.
1.4 A within-laboratory detection estimate is useful in characterizing the concentration below which a method, for an analyte, as implemented in a specific laboratory, does not (with high confidence) discriminate the presence of the analyte from that of the absence of an analyte. As such an estimator, the WDE Standard (and the WDE and WCL values produced through its application) are useful where a trace-analysis testing method needs to be used.
5.1 This practice can be used in a single laboratory for trace analysis (that is, where: 1) there are concentrations near the lower limit of the method and 2) the measurements system’s capability to discriminate analyte presence from analyte absence is of interest). In these testing situations, a reliable estimate of the minimum level at which there is confidence that detection of the analyte by the method represents true presence of the analyte in the sample is key.
Where within-laboratory detection is important to data use, the WDE procedure should be used to establish the within-laboratory detection capability for each unique application of a method.
5.2 When properly applied, the WDE procedure ensures that the 99 %/95 % WDE has the following properties:
Note 1—The referenced probabilities, α and β, are key parameters for risk-based assessment of a detection limit.
5.3 When this practice is utilized by a laboratory to develop these false-positive- and false-negative-control point estimates, from data representative of routine operations, the laboratory may confidently claim these levels of false-positive and false-negative control in the future, so long as the data used remain representative of that future operation.
The laboratory may also qualify reported data using the appropriate point estimates (for example YC, YD, WCL, WDE) or censor data below the WCL as a valid basis for these data-reporting practices.
5.4 The WDE practice applies to measurement methods for which calibration error (that is, the error in the calibration of the measurement system) is minor relative to the combined other sources of variability. Some examples of other sources and when they may be dominant are:
5.5 Reducing calibration error by use of allowable, though more stringent, calibration procedures (for example, multiple concentrations, replication, tight calibration-acceptance criteria, etc.) and through calibration verification (for example, analysis of a traceable standard from a second, independent source, calibration diagnostics) can reduce the magnitude of the calibration error.
5.6 Alternative Data-Quality Objectives—Other values for α, β, confidence, etc. may be chosen as parameters; however, this procedure addresses only those stated here in.
5.7 Collectively, the many sources of variation combine to cause within-laboratory measurements at any true concentration to be normally distributed. The assumption of normality is important for some of the statistics used; data normality should be assessed if there is reason to believe this assumption is not valid.
5.8 If control of false negatives is not a data-quality objective, the WCL determined through this procedure provides a sound criterion for future determination of false-positive control; in such cases, the laboratory may confidently claim that true values above the WCL have a statistically significant difference from like-matrix zero-concentration samples (for example, from the method blank), but nothing more.
5.9 Where as-measured values (for example, not corrected for bias), not true values are of interest, YC and YD may be used as these as-measured levels of the WCL and WDE.