Standard Test Method for Analytical Procedure for the Analysis of Ultrashort Chain Perfluoroalkyl Substances (PFAS) in Aqueous Samples by Liquid Chromatography/Mass Spectrometry (LC-MS/MS)
1.1 This test method covers the analysis of ultrashort chain perfluoroalkyl substances (PFAS) listed in Table 2, in aqueous samples. Ultrashort chain PFAS are defined as PFAS with a carbon chain length of three or fewer.
After spiking with isotopically labeled extracted internal standards (EIS) and cleanup on solid phase extraction (SPE) cartridges, aqueous samples are analyzed by liquid chromatography/mass spectrometry (LC-MS/MS). Final sample concentrations are determined by isotope dilution or surrogate standard quantification using the EIS.
1.2 Samples may be extracted and analyzed for ultrashort chain PFAS only or, where subsample size requirements permit, they may also be extracted for longer chain PFAS at the same time.
1.3 Units—The values stated in SI units are to be regarded as the standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
5.1 PFAS are widely used in various industrial and commercial products; they are persistent, bio-accumulative, and ubiquitous in the environment. PFAS have been reported to exhibit developmental toxicity, hepatotoxicity, immunotoxicity, and hormone disturbance.
PFAS have been detected in soils, sludges, surface, and drinking waters. This is a quick, easy, and robust method to determine these compounds quantitatively at trace levels in water matrices.
5.2 This test method has been validated using reagent water. Refer to Section 15.