ASTM E1045-00(2022)

Current Revision
Standard Specification for Pipet, Sahli Hemoglobin

Best Price Guarantee Jan 1, 2022 English 3 Pages

Sub Total (1 Item(s)) $ 0.00
Estimated Shipping $ 0.00
Total (Pre-Tax) $ 0.00

You own the digital copy of this edition

View in Library
or
ASTM E1045-00(2022) PDF Preview
ASTM E1045-00(2022)

This specification covers reusable Sahli hemoglobin pipets.

The pipets shall be made of common spirit bore white back tubing. Pipet design shall be straight and of one-piece construction. Pipet delivery tips shall be made with a gradual or concave taper. Pipet dimensions shall be within tolerance limit indicated in this specification.

The pipet shall have markings which include:

  • graduation line
  • volumetric designation
  • identification
  • capacity deviation

Testing methods include capacity test, capacity deviation, pigmentation test, and strain free test.

1.1Β This specification covers reusable pipets calibrated β€œto contain” 20 cmm of whole blood and used for hemoglobin determinations.

1.2Β The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3Β This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.