Electroacoustics - Sound calibrators
Electroacoustics - Sound calibrators
IEC 60942:2017 specifies the performance requirements for three classes of sound calibrator: class LS (Laboratory Standard), class 1 and class 2. Acceptance limits are smallest for class LS and greatest for class 2 instruments.
Class LS sound calibrators are normally used only in the laboratory; class 1 and class 2 are considered as sound calibrators for field use. A class 1 sound calibrator is primarily intended for use with a class 1 sound level meter and a class 2 sound calibrator primarily with a class 2 sound level meter, as specified in IEC 61672-1.
The acceptance limits for class LS sound calibrators are based on the use of a laboratory standard microphone, as specified in IEC 61094-1, for demonstrations of conformance to the requirements of this document. The acceptance limits for class 1 and class 2 sound calibrators are based on the use of a working standard microphone, as specified in IEC 61094-4, for demonstrations of conformance to the requirements of this document.
This fourth edition cancels and replaces the third edition published in 2003, of which it constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
Failed to load document history.
| Publish Date | Document Id | Type | View |
|---|